Specification:-
Product Vitalograph Pneumotrac
Model Number 6800
PC Operating System Networked or Stand Alone Windows 7, 8, 8.1, 10
Size 103mm x 185mm
Parameters Selectable
Operating Temperature Range 17 - 37°C Design Limits 10-40°C
Power Supply From USB port
Flow Detection Principle Fleisch type pneumotachograph
Volume Detection Flow integration sampling @ 100Hz
Accuracy when in Operating Range Volumes: Better than +/- 3% (Max 8L / Min 0L) Flows: Better
than +/- 5%
(Max 16L/s / Min 0.02L/s)
Linearity: +/- 1% in range 0.1 L/s to 16 L/s
Max Test Duration 45 seconds
Performance Standards ATS/ERS 2019
EN ISO 23747:2015
Safety Standards IEC 60601-1:2006 + A1:2013
Medical Safety Standard Medical Devices Directive 93/42/EEC
Designed & Manufactured Under ISO 13485:2003 and 2016
FDA 21 CFR 820
CMDR
The special design of the Base Unit allows the Flowhead to be kept in an inclined position
to remove trapped liquid particles inside the Flowhead, if any.